The biochemical incubator may encounter some minor faults during use. If you encounter a fault, you can check the accompanying instructions first. The general manual contains the biochemical incubator maintenance method .

The following is a description of the troubleshooting of biochemical culture and biochemical incubator maintenance methods for Shanghai Dongsheng Instruments :

Fault phenomenon

Reason for failure

Troubleshooting

No power at boot

(Light does not shine)

The power outlet has no power or has poor contact with the plug

repair

Box power cord is broken or plug is not plugged in

Repair, reinsert

Power switch is broken (or not open)

Replace and open the power switch

Fuse off

If the power is still blown after the change, it is necessary to check whether the motor and other components are short-circuited (the motor has bad odor, the wire package is black, etc.) or the through-shell (insulation resistance is 0).

Meter no display

Wiring off or instrument burnout

Reconnect the wires or replace the meter

the meter shows

"- - - -"

The sensor is broken or the wiring is broken (falling off)

Repair or replace Pt100 (100Ω, 0.3Ω/°C at 0°C)

Temperature value is out of control range

Use the appropriate temperature value within the specified value

Not cooling

The switch is not in the correct position

Set the correct position

Compressor is broken or pipeline blockage or R12 leak

Replace, block, add R12

Compressor overheat protector start

When the compressor temperature drops, it naturally recovers

Compressor protector burned out (burnt smell)

Replacement protector

The temperature is out of control and

AL1 alarm light is on

AL1 setting is incorrect

Reset

Two-way thyristor is broken down

Temperature controller is bad (AL1 alarm light is off)

Replace the temperature controller

Not warming up

Large temperature control error

Or have a static difference

Temperature controller is bad

HEAT light bright two-way thyristor non-conducting

Replace the temperature controller

The timing function is used, and the temperature rises and then falls.

Make T1≠0

Not warming up

Large temperature control error

Or have a static difference

The fan is broken (not turning)

Replace the fan

Pt100 has poor contact and the resistance becomes larger

Rewiring

Parameter (HL, SC) setting is incorrect

Reset

PID and other parameters are set incorrectly

Re-adjust parameters

The condenser has too much frosting and low cooling power

Warming down

Fan sound is abnormal

Or noisy

(greater than 70db(A))

The fan is broken or the compressor fan is broken.

Replace the fan

Rub the rear duct plate or the fan blade

Repair, replacement

If abnormalities, smells, smog, etc. cannot be handled during use, please turn off the power immediately. Users should not blindly repair. They should inform Shanghai Dongsheng Instrument Equipment Co., Ltd., and the repairs are provided by professionals. We provide home repair service.

200W Laser Cleaning Machine

We produce and export full series of Laser Cleaning Machine, this laser cleaner is handheld and manual, which is quite convenient for industrial use. It can clean tyre mold, wheel hub, remove rust, oil, paint or deal with welding seams or oxide layer etc.

raycus laser cleaning machine, industrial laser cleaner, industrial laser cleaning machine

Feiquan Laser Technology Wuxi Co., Ltd. , https://www.fq-lasers.com